Detector Adjustments

Temperature Correction Factor (CF1)*

Part Number 39°F 50°F 68°F 84°F 86°F 104°F Usable Range of
Relative Humidity
 
CO2-0.25–3.0% 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
CO2-0.5–10.0% 0.9 1.0 1.0 1.0 1.1 1.2 10 – 90%  
CO2-1–20% 0.9 1.0 1.0 1.0 1.1 1.2 10 – 90%  
CO2-5–60% 0.9 1.0 1.0 1.0 1.1 1.2    
CO2-150–2500PPM 0.9 1.0 1.0 1.0 1.1 1.2 10 – 90%  
H2O-0.1–2.0MGL     (refer to spec sheet)       N/A  
H2O-1–18MGL     (refer to spec sheet)       N/A  
H2O-1–30MGL 1.0 1.0 1.0 1.0 1.0 1.0 N/A  
H2O-2–10LBMMCF   1.3 1   0.7 0.5 N/A  
H2O-6–40LBMMCF   1.2 1   0.7 0.5 N/A  
H2S-0.1–2.0% 1.0 1.0 1.0 1.0 1.0 1.0    
H2S-0.2–7.0% 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
H2S-0.5–4.0% 1.0 1.0 1.0 1.0 1.0 1.0    
H2S-0.5–10.0% 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
H2S-2–40% 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
H2S-1–30PPM 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
H2S-2.5–60PPM 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
H2S-5–100PPM 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
H2S-10–250PPM 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
H2S-50–800PPM 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
H2S-100–2000PPM 1.0 1.0 1.0 1.0 1.0 1.0 10 – 90%  
PH3-0.1–10PPM 1.0 1.0 1.0 1.0 1.0 1.0 40 – 60%  

*Multiply the observed reading by the correction factor (CFV * CF1) to obtain the truc concentration.